Small Business / Woman Owned, HUB Certified - State of Texas

23230 Sandsage Lane
Katy, Texas 77494-4207
Phone: (281) 394-2311
Fax:     (281) 394-2522

Medical Instruments
  Diagnostic Radiology
  Dosimetry
  Hand Held Survey Meters
  Health Physics
  Lead Aprons
  Mammography
  Medical Physics
  Nuclear Medicine
  Radiation Safety
  Radiation Therapy
  Radiation Survey Meters
  Tables and Accessories
  Ultrasound
  X-Ray Test Devices
Homeland Security
  Area Monitors
  Response Kits
  Hand Held Survey Meters
  Portal Monitors
  Waste Monitors
  Gate Monitors
  Personal Dosimeters
Special Equipment
  Calibration Services
  Scintillation Detectors
Continuing Education
  Nursing
  Radiology
Software
  Lead Apron Inspection


Medical Physics : CR RADCHEX
CR RADCHEX Model 07-CRX

  • Calibrate computed radiography (CR) plate readers and automatic exposure control (AEC)
  • Assess ongoing performance of the CR plate reader, AEC and automatic programmed radiography (APR)
  • Set and maintain desired clinical system speed (dose) of the CR system
  • CR RADCHEX is an indispensable tool for service engineers, physicists and quality assurance personnel
Introduction
The CR RADCHEX System, consisting of a 24 x 30 cm electronic cassette, electronic interface box, universal calibration tool, and Excel software, can assess the performance and proper calibration of computed radiography (CR) automatic exposure control (AEC) systems. CR RADCHEX can be used to perform periodic assessment. The system provides information to make recommendations for corrective action and adjustments of clinical system speed and thus dose to patients. CR plate readers, AEC and automatic programmed radiography (APR) can all be correlated to maximize the function of computed radiography systems in any clinical setting.

Applications
CR RADCHEX is ideal for use by service engineers to initially calibrate and troubleshoot the CR plate reader, AEC and density selector settings. Physicists use CR RADCHEX to assess the performance of CR-AEC for compliance to clinical system speed objectives and patient dose. QA personnel can use CR RADCHEX to periodically document the performance of the CR system and to compare CR to film/screen systems regarding desired ALARA objectives. CR RADCHEX is designed to work with all major brands of CR equipment.

Features
The electronic cassette once calibrated produces an output signal directly proportional to the light output from an exposed CR imaging plate when the plate is stimulated by laser light (tracking ± 3% for energy and patient density). An interface box converts the output signal to a digital number displayed on a LCD (CR-AEC number) that further corresponds to a relative clinical system speed number also displayed on the LCD.

Excel software analyzes input (CR-AEC number) and compares it to the imaging plate signal output number (SAL# AGFA, EI# Kodak, S# Fuji, Rex# Konica). Three templates are included for service engineers, medical physicists/radiation protection inspectors and QC personnel.
 


Site created & maintained by
Tux Studios.Com

© Owens Scientific, Inc., Houston,Texas


Advanced Search

New Items
Unfors DXR+
Direct X-ray Ruler
 
High Sensitivity
Personal Radiation Detector
 
Inovision Model 451B
Ion Chamber Survey Meter with Beta Slide
 
Frustrated with lead apron inspections?
Easily Enter, Organize and Report your Lead Apron Tracking Information